Performing and Reviewing Calculations
Trends and Calculations   8-23
Tr
e
nds
 a
nd 
C
a
lcula
ti
o
n
s
  
Notes—
If you plan to use monitored parameter values in Calculations, be 
sure that when you are configuring parameters, you assign the 
parameter the same label that is used in Calculations. For example, 
ABP should be used instead of ART because ABP is the label used 
by Calculations. 
When you enter the Task Window, the Calculation time will be the 
last C.O. time or the current time, depending on how the monitor is 
configured. If C.O. time is used, values stored more than one minute 
before that time, are automatically entered. Values up to and 
including 30 minutes before the calculation time will be used. The 
most recent values occurring before the calculation time will be 
used for height, weight, and head circumference calculations.
 Displayed values may have been automatically stored or manually 
entered using the keypad or 
You must verify that all input values are appropriate for your patient 
prior to performing calculations.
If you enter pressure values manually, you must enter all three parts 
- systolic, diastolic, and mean - in order for calculations to be 
performed for that pressure reading. 
BSA Formula
The patient monitors provide both the Boyd and Dubois formulas for the 
calculation of body surface area (BSA). The default configuration for all 
patients is the Boyd formula. All calculation results that use BSA are 
indexed to the selected formula. The Hemodynamics and Oxygenation 
Calculations Task Windows which display BSA use the label BSA(B) for 
the Boyd formula and BSA(D) for Dubois. To change the default [for 
example to use Dubois for adults and keep Boyd for neonates], the 
appropriate selection must be made in the Patient Data Configuration 
Task Window. See your Philips Service Engineer or biomedical 
engineering department for assistance in making a change.
Resample Vitals
.