Philips V26CT 用户手册

下载
页码 392
Adjusting Neonatal Event Review Settings
Neonatal Event Review   9-19
Ne
ona
ta
l Ev
e
n
Revi
ew
  
The Event Setup task window enables you to adjust the following items: 
Event 
Criteria
The event type (Apnea, Bradycardia, Desaturation) determines how a 
specific event is triggered and stored on the monitor.
Apnea Event
Settings
Apnea events are always triggered by a Red*** Apnea alarm. The Apnea 
alarm is derived from the RESP parameter and occurs whenever 
breathing ceases for a period exceeding the defined Apnea alarm limit 
(usually 20 seconds).
The Apnea alarm limit and trigger time can be viewed but not adjusted in 
the Event Setup task window. These settings must be adjusted in 
Monitoring Mode
, in the RESP task window. 
Trigger Mode
The trigger mode determines whether an 
event is triggered by a physiological alarm or 
by criteria defined specifically by the user.
Trig Threshold
The trigger threshold can be adjusted if the 
user defined trigger mode setting has been 
selected
Trigger Time
The trigger time can be adjusted if the user 
defined trigger mode setting has been 
selected.