Adjusting Neonatal Event Review Settings
9-20     Neonatal Event Review  
Ne
ona
ta
l Ev
e
n
Revi
ew
  
Bradycardia
Event Settings
Alarm Triggered
The Bradycardia event is triggered by a Red*** Brady alarm.
The Bradycardia alarm limit can be viewed but not adjusted in the Event 
Setup task window. The alarm limit must be adjusted in Monitoring 
Mode
 in the HR/PULSE or Alarm Limits task window and the trigger 
time must be adjusted in Configuration Mode in the HR/PULSE task 
window. 
If the Bradycardia alarm related trigger threshold and trigger time items 
are selected, the 
 
key is inactive and the following 
status message is displayed in the Event Setup task window:
“Brady items must be changed in the HR/PULSE Setup”
User Defined
For situations where the storage and documentation of an event should 
be independent of all Bradycardia alarms, you can adjust the 
Bradycardia event threshold and trigger time in Monitoring Mode in 
the Event Setup task window.
Note—
When the patient monitor is connected to an Arrhythmia System, 
the Bradycardia alarms are generated by the latter and not the former. 
The alarm limits can only be viewed and adjusted on the Arrhythmia. In 
this situation, the Bradycardia threshold line is displayed in the Event 
Review task window as a shaded area and the Brady trigger time and 
threshold items are set to “From Arrhyth” in the Event Setup task 
window.