Data Transfer Module
Data Transfer   10-5
D
a
ta
 T
ra
n
sf
er
  
When the module is unplugged, it retains data for at least one hour.
After data is transferred to the monitor, the module is automatically 
erased. This is to ensure that the module is empty when it is used for the 
next transfer.
Symbols to 
Indicate Key 
Functions
As detailed in the table below, some versions of this module now use 
symbols, instead of words, to indicate the function of some keys. If the 
monitor’s Reference Manual tells you, for example, to select the Monitor 
key, you should press the key marked with the corresponding “monitor” 
symbol. 
The design change also means that you will now find the module’s 
product number on the rear of the module’s housing, not on the front. 
What is 
Transferred
The module is capable of storing and transferring up to 24 hours of trend 
data and demographic information. When a transfer is initiated, the 
following data is transferred: 
Continuously monitored parameters (for example, ECG) 
Resampled Values
Aperiodic measurements (for example, CO) 
All patient demographic information except bed number 
The time at which calculations were performed in the 
Hemodynamic, Oxygenation, and Ventilation calculation task 
windows 
Marked events 
ST events (lead changes, reference changes) 
The following data is not transferred:
Waveforms, including ST and CO waveforms 
Split Screen trends 
Arrhythmia events 
Symbol
Name
Function
Which Modules?
Monitor
initiates transfer from module to 
monitor 
Data Transfer Module (M1235A)
Module
initiates transfer from monitor to 
module