Data Transfer Module
10-6     Data Transfer  
D
a
ta
 T
ra
n
sf
er
  
The order in which the parameters are listed on the display is relative to 
the destination monitor, not the originating monitor.
Patient demographics have the highest priority in the order of data being 
transferred. The aperiodic parameters are transferred next, before 
continuously monitored parameters, for each hour. During a transfer to 
the monitor, aperiodic data obtained by the receiving monitor after the 
start of the transfer will have higher priority than the aperiodic data 
from the module.
Note—
If the module is unplugged before a transfer is complete, some 
data may not be transferred. 
With the exception of gas parameters, if a parameter has different units 
at the source and destination monitors, the data from the source will be 
converted to the units of the destination. 
For gas parameters, there is no unit conversion between volume% and 
pressure (mmHg/kPa), and vice versa. To transfer gas parameters using 
a Data Transfer Module, the gas units at the source monitor must match 
the default units at the destination monitor. So, for example, if the CO
2
 
units are configured for volume% at the source and for mmHg at the 
destination, ETCO
2
 and IMCO
2
 values will not be transferred.
If the monitor is put in Monitor Standby during a transfer, all processing 
in the monitor will stop, but the transfer will continue until it has 
completed, or the module is unplugged. The module keys will be 
disabled in Monitor Standby. When the monitor comes out of standby, 
regular updates to the module will continue at a rate of once per minute.
Transfers can occur in both Monitoring and Demo modes. Transfers 
between modes can only be made from a module in Monitoring Mode to 
a monitor in Demo mode.
Resampled and other manually entered data is annotated with an 
asterisk (*) at the destination monitor only if both source and 
destination monitors have software for Release F or higher (CMS) or 
Release A.0 or higher (V24 and V26).