Philips V26CT 用户手册

下载
页码 392
Vital Signs, Blood Review and Graphs
Data Transfer   10-17
D
a
ta
 T
ra
n
sf
er
  
Vital Signs, Blood Review and Graphs 
Some functions under Trends and Procedures are unavailable during 
transfers of patient data. 
Caution
Transferred data is not entered into the module database until up to one 
minute after it enters the monitoring system. Because of this delay, during 
updates to the module, some aperiodic data may not be transferred for up 
to 2 minutes after the actual time it was obtained. 
Time Stamp
Transferred data is identified in Patient Data displays by a “T” placed 
before the time in the Vital Signs, Graph Trends, and Calculation Review 
task windows. Transferred data is labeled relative to the time in the 
destination monitor. The Data Transfer module has a clock which ages 
the data the exact amount of time which has transpired between when the 
measurement was obtained and when the data is put into the receiving 
monitor.
To the Monitor
 
or
 
To the Module
 
or
 
Vital Signs
Pts
Refernce
Blood Review
Pts
Refernce
Graph Trends
Pts
Refernce
Discharg Patient
Pts
Refernce
End Case
Pts
Refernce
ST Analysis
Pts
Refernce
Discharg Patient
Pts
Refernce
End Case
Pts