Troubleshooting
Data Transfer   10-19
D
a
ta
 T
ra
n
sf
er
  
Troubleshooting
Caution
In a centrally networked system, if the previous patient is discharged only 
at the monitor and a new patient's data is transferred to the monitor, the 
new patient's name and ID will be overwritten by the central station's 
broadcast of the previous patient's name. To prevent this situation from 
occurring, the previous patient should be discharged from central station 
before
 the transfer. 
The following table outlines the Data Transfer behavior under various 
events.
Event
Results
Power Off < 1 min
Transfer will restart automatically
Power Off > 1 min
Transfer is stopped. Collection of new data 
must be restarted manually.
Op. Mode Change
Transfer is stopped. Data in module remains 
intact.
Config Mode Change
Transfer is stopped. Data in module remains 
intact.
Patient Discharge or End 
Case
Transfer is stopped. Data in monitor is 
deleted. Data in module remains intact.