Performance Specifications
10-20     Data Transfer  
D
a
ta
 T
ra
n
sf
er
  
Performance Specifications
For safety and environmental specifications, please refer to Chapter 11, 
“Monitor Installation and Patient Safety”. 
Data 
Transfer 
Module
Transfer Time 
A typical transfer time for 24 hours of data is under 30 seconds. The 
transfer time could be as high as 60 seconds for highly customized 
trends.
Back up Memory Time 
The Data Transfer Module will store data for a minimum of 1 hour under 
the following conditions: 
 the module is out of the monitor, 
 the module is in the rack of a powered Off monitor, 
 the monitor has a power failure.