Monitor Maintenance
Maintenance   13-7
M
a
int
e
na
nc
e
  
Preventative Maintenance   
Replace backlight tube 
assembly   
Cleaning procedures
2.5 years of continuous use 
or as needed 
Perform the Preventive 
Maintenance Procedure for 
the M1180A #A01/A02 Wall 
Mount Assembly every year.
See “Testing and 
Maintaining the CMS” in the 
“CMS Service Guide” or 
“Testing and Maintaining 
the V24 and V26“in the “V24 
and V26 Service Guide”.
Full performance checks 
including safety checks.
At least every two  years,  
after any repairs, or as 
needed.
See “Testing and 
Maintaining the CMS” in the 
“CMS Service Guide” or 
“Testing and Maintaining 
the V24 and V26“in the “V24 
and V26 Service Guide”.
Check Calibration of NBP & 
Sidestream CO
2
Once a year
See chapter 3 of the CMS 
Service Guide “Maintaining 
the CMS”.
Check Calibration of all 
parameters except NBP & 
Sidestream CO
2
Every 2 years
See chapter 3 of the CMS 
Service Guide “Maintaining 
the CMS”.
Procedure
Frequency 
Source of Information