Monitor Maintenance
13-8     Maintenance  
M
a
int
e
na
nc
e
  
Inspect the 
System
The following lists what to look for when inspecting the system:
Inspect the
Exterior of the
Monitor
1. Examine the exterior of the unit for cleanliness and general 
physical condition. Ensure that plastic housings are intact, that all 
hardware is present and tight, and that there are no spilled liquids 
or other serious abuse. 
2. Inspect connectors of the Rack and ensure that the Modules are 
locked into place and do not slide out without first releasing the 
locking plate on the bottom of the Module. 
3. Inspect all accessories external to the system such as transducers, 
referring to the manufacturer's documentation. 
4. Verify the backlight tube brightness is adequate. If the brightness 
is not adequate, notify your biomedical department or the Philips 
representative. Philips recommends replacement of the backlight 
tube every 18 months of continuous use. 
Inspect the
Cables and
Cords
1. Examine the line/power plug for damage. Ensure that the prongs 
of the plug do not move in the casing. 
2. Inspect the line/power cord for signs of damage. If damaged, 
replace the entire cord. Ensure that any new line/power cord or 
plug is fitted with the same polarity as the old one. 
3. Inspect the interconnecting cable between the module rack and 
the mainframe for general condition. Examine it carefully to detect 
breaks in the insulation. Ensure that the cable connectors are 
properly engaged and that any connector screws are fully 
tightened. 
4. Inspect the patient cables and leads and their strain reliefs for 
general condition. Examine cables carefully to detect breaks in the 
insulation and to ensure that they are gripped securely in the 
connectors of each end to prevent rotation or other strain. 
5. Flex the patient cable near each end to verify that there are no 
intermittent faults.