Monitor Maintenance
13-10     Maintenance  
M
a
int
e
na
nc
e
  
Perform a 
System Self-
Test
The following lists how to perform the system self-test.
1. Examine all controls for physical condition. During the course of 
this examination, ensure that the controls perform their proper 
function 
2. Check that the parameter module LED's light up when first 
plugged in. 
3. Confirm that the display lights up and displays a Main Screen. 
4. Press the ECG key on the front of the ECG module in a system 
that is turned On. This brings up the ECG Task Window and 
verifies that the System board and integral rack are 
communicating with the plug-in module. If the ECG module does 
not respond, perform the ECG module Functional self-test. 
5. Ensure correct operation (visually and audibly) of all alarms using, 
if required, the User Guide and a patient simulator. 
6. If possible connect the ECG parameter module of the unit to an 
ECG simulator and verify that an adequate trace is received at 
each patient lead selection. 
7. Press 
. In this 
screen, press 
. This provides a realtime list of the 
modules in the rack. Remove a module and observe the list 
changes to delete the module you removed. Re-insert the module 
and observe the list changes to include the module. 