Performance Assurance Checks
Maintenance   13-13
M
a
int
e
na
nc
e
  
Functional 
Testing 
Procedures
The system has a module self-test which generates and displays test 
waveforms and corresponding numerics for all the modules that are 
connected and turned on. The test signals are displayed for about 30 
seconds and then the display returns to monitoring mode. The test signal 
will be displayed only if the system self-test was successful and the 
parameter is plugged into the system and turned on. 
Perform the following quick system check to verify that System board, 
integral rack and modules are communication properly before verifying 
the individual parameter modules.
Perform a
Quick System
Check
You can verify that the System board, integral rack and modules are 
communicating properly by completing the following test.
TEST - Press the ECG key on the front of the ECG module in a system 
that is turned On.
RESULT - Pressing the ECG key brings up the ECG Task Window and 
verifies that the System board and integral rack are communicating with 
the plug-in module.
If the plug-in module Task Window, waveform or numeric is missing from 
the display, the module may not be turned On. Try pushing the key on the 
plug-in to see if the system is communicating with the module.
If the ECG module does not respond, perform the module self-test.
To turn a parameter On, press the following key sequence:
1. Press  
2.
3.
4.
Module Setup
Parametr On/Off
Pts
Refernce
Select Parametr
Pts
Refernce
On/Off Setup 
Pts
Refernce