Performance Assurance Checks
13-14     Maintenance  
M
a
int
e
na
nc
e
  
Procedure
Overview
1. Choose a patient category (Adult/Pedi/Neo) by pressing:
 
 
 
 
2. Display 30 seconds of the waveforms and numerics of all 
connected modules by pressing:
 
 
3. Repeat Step 2 as necessary to check the individual test waveforms 
and numerics for the category. 
4. Use Step 1 to choose a new patient category and repeat Steps 2 
and 3. 
See the following table for test values.
System Self-test Values 
Monitor Setup
Config Sets
Pts
Refernce
Change Pat Categ
Pts
Refernce
Monitor Setup
Test Signals
Pts
Refernce
Signals
Module 
Test Numeric 
Test Waveform
ECG and ECG/
RESP
ECG
100 bpm in ADULT mode 
125bpm in NEO/PED mode
Simulated Square Wave 
RESP
15 rpm in ADULT mode 
30 nrpm in PEDI mode 
55 rpm in NEO mode
Simulated Wave
Pressure
Simulated square wave 
Systolic
120 mmHg ADULT 
60 mmHg PEDI/NEO
Diastolic
0 mmHg ADULT 
0 mmHg PEDI/NEO
SpO
2
/Pleth
SpO
2
100%
No Wave   
Pleth
Wave on display
 Simulated wave