Chapter 6
Performing Sweep Measurements
6-2
ni.com
generator, function generator, dynamic signal analyses, or 
multifunction I/O (MIO) device.
3.
Click the Add Step button, shown at left, and select Execution 
Control»Sweep
.
4.
Click the Add button in the Step Setup dialog box to display the list 
of sweep parameters from each step in the project, as shown in 
Figure 6-1.
Figure 6-1.  Sweep Parameter Selection Dialog Box
5.
Select the Frequency parameter under Create Signal and click the 
OK button.
The Sweep step encloses the Create Signal step, which provides the 
signal to sweep.
6.
On the Sweep Configuration tab, select Exponential from the Type 
pull-down menu.
7.
Enter 
1k
 in the Start Frequency (Hz) field, and enter 
40k
 in the Stop 
Frequency (Hz) field.
8.
Enter 
150
 in the Number of points field.
The Sweep Configuration tab should appear as shown in Figure 6-2.