Sources
Internal, Onboard Clock (internal VCXO)*
External, CLK IN (front panel SMB connector)
*
Internal Sample
Clock is locked to the
Reference Clock or
derived from the
onboard VCXO
Reference Clock or
derived from the
onboard VCXO
Onboard Clock (Internal VCXO)
Sample Rate
Range
Range
Real-Time Sampling
(Single Shot)
Random Interleaved
Sampling (RIS)
*
Divide by n
decimation used
for all rates less than
250 MS/s
for all rates less than
250 MS/s
For more information
about Sample Clock
and decimation, refer
to the NI High-Speed
Digitizers Help.
about Sample Clock
and decimation, refer
to the NI High-Speed
Digitizers Help.
3.815 kS/s to 250MS/s*
250 MS/s to 5 GS/s in
increments of 250 MS/s
increments of 250 MS/s
Timebase
Frequency
Frequency
250 MHz
When not using
External Sample
Clock
External Sample
Clock
Timebase
Accuracy
Accuracy
Not Phase-Locked to
Reference Clock
Phase-Locked to
Reference Clock
Reference Clock
—
±25 ppm
Equal to the Reference
Clock accuracy
Clock accuracy
Sample Clock
Delay Range
Delay Range
±1 Sample Clock period
—
Sample Clock
Delay Resolution
Delay Resolution
≤20 ps
—