National Instruments NI PCI-5114 用户手册

下载
页码 23
NI PXI/PCI-5114 Specifications
8
ni.com
CLK IN (Sample Clock and Reference Clock Input,
Front Panel Connector)
Trigger
Reference (Stop) Trigger
Specification
Value
Input Voltage Range
Sine wave: 0.65 V
pk-pk
 to 2.8 V
pk-pk
 (0 dBm to 13 dBm)
Square wave: 0.2 V
pk-pk
 to 2.8 V
pk-pk
Maximum Input 
Overload
7 V
rms
 with |Peaks| 
≤10 V
Impedance
50 
Ω
Coupling
AC
Specification
Value
Comments
Trigger Types 
and Sources
Types
Sources
Refer to the 
following 
sections and to 
NI High-Speed 
Digitizers Help
 
for more 
information. 
Edge, Window, Hysteresis, 
Video, Digital, Immediate, 
and Software
CH 0, CH 1, TRIG, 
PXI_Trig<0..6>, PFI <0..1>, 
PXI Star Trigger, 
RTSI<0..6>, and Software
Time 
Resolution
TDC
Onboard Clock
External Clock
TDC = Time to 
Digital 
Conversion 
Circuit
On
40 ps
N/A
Off
4 ns
External Clock Period
Rearm Time
TDC
Rearm Time
Holdoff set to 0
On
10
μs
Off
2
μs
Holdoff
From Rearm Time up to [(2
35
 – 1) 
× (Sample Clock Period)]
Trigger Delay
From 0 up to [(2
35
 – 1) – posttrigger samples
× 
(1/sample rate), in seconds