National Instruments NI 4070/4072 6 1/2-Digit FlexDMM 用户手册

下载
页码 68
NI 4070/4072 Calibration Procedure
10
ni.com
Figure 1.  Cable Connections for Voltage and 2-Wire Resistance
23. Wait two minutes for the thermal EMF to stabilize.
24. Generate 0 V on the calibrator. 
25. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_VOLTS
Range = 
0.1
Resolution = 
100e–9
26. Set the input resistance of the NI 4070/4072 to >10 G
Ω by calling 
niDMM_SetAttributeViReal64
 with the following parameters:
Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = 
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
 
1
NI 4070/4072
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cable
Table 1.  Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4070/4072)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post Label
(Fluke 5700A/5720A Calibrator)
HI
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD
HI
LO
300V
MAX
HI
LO
AUX
I/O
1A, 250V
MAX
300V
MAX
5V
MAX
AMPS
INPUT 
SENSE
W 4W
6½-Digit FlexDMM
CAT II
 
V
W
HI
LO
HI
LO
HI
SENSE
V
Ω
OUTPUT
V
ΩA
AUX
CURRENT
GUARD
GROUND
1
2
3