NI 4070/4072 Calibration Procedure
8
ni.com
Verifying DC Voltage 
To verify DC voltage of the NI 4070/4072, complete the following steps:
1.
Plug in the insulated banana plug shorting bar across the HI and LO 
banana plug connectors on the NI 4070/4072.
2.
Wait one minute for the thermal EMF to stabilize.
3.
Call 
niDMM_reset
.
4.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_VOLTS
Range = 
1
Resolution = 
1e–6
5.
Set the input resistance of the NI 4070/4072 to >10 G
Ω by calling 
niDMM_SetAttributeViReal64
 with the following parameters:
Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = 
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
 
6.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 15.
7.
Set the input resistance of the NI 4070/4072 to 10 M
Ω by calling 
niDMM_SetAttributeViReal64
 with the following parameters:
Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = 
NIDMM_VAL_10_MEGAOHM
 
8.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 15.
9.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_VOLTS
Range = 
10
Resolution = 
10e–6
10. Set the input resistance of the NI 4070/4072 to >10 G
Ω by calling 
niDMM_SetAttributeViReal64
 with the following parameters:
Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = 
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
 
11. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 15.