NI 4070/4072 Calibration Procedure
18
ni.com
c.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
d.
Call 
niDMM_ConfigureMeasurement
 again, changing Mode to 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
e.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
18. Output 5 V at 30 Hz on the calibrator.
19. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 
5
Resolution = 
5e–6
20. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
Table 3.  niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output
niDMM_ConfigureMeasurement Parameters
 Amplitude
Frequency
Function
Range
Resolution
1
500 mV
50 Hz 
NIDMM_VAL_AC_VOLTS
0.5
500e–9
500 mV 
50 Hz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
500e–9
2
500 mV 
1 kHz
NIDMM_VAL_AC_VOLTS
0.5
500e–9
500 mV 
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
500e–9
3
500 mV 
1 kHz
NIDMM_VAL_AC_VOLTS
5
5e–6
500 mV 
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
5
5e–6
4
500 mV 
20 kHz
NIDMM_VAL_AC_VOLTS
0.5
500e–9
500 mV 
20 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
500e–9
5
500 mV
50 kHz
NIDMM_VAL_AC_VOLTS
0.5
500e–9
500 mV
50 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
500e–9
6
500 mV
100 kHz
NIDMM_VAL_AC_VOLTS
0.5
500e–9
500 mV 
100 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
500e–9
7
500 mV 
300 kHz
NIDMM_VAL_AC_VOLTS
0.5
500e–9
500 mV
300 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
500e–9