NI 4070/4072 Calibration Procedure
20
ni.com
23. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 
50
Resolution = 
50e–6
24. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
25. Refer to Table for the appropriate calibrator outputs and parameter 
values as you complete the following steps:
a.
On the calibrator, output the value listed in the Calibrator Output 
column in Table 5 for the current iteration.
b.
Call 
niDMM_ConfigureMeasurement
 with Mode set to 
NIDMM_VAL_AC_VOLTS
 and the remaining parameters as shown 
in Table 5 for the current iteration. 
c.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
d.
Call 
niDMM_ConfigureMeasurement
 again, changing Mode to 
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
e.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 16.
Table 5.   niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output
niDMM_ConfigureMeasurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
50 V 
50 Hz
NIDMM_VAL_AC_VOLTS
50
50e–6
50 V 
50 Hz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
50e–6
2
50 V 
1 kHz
NIDMM_VAL_AC_VOLTS
50
50e–6
50 V 
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
50e–6
3
50 V 
20 kHz
NIDMM_VAL_AC_VOLTS
50
50e–6
50 V 
20 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
50e–6
4
50 V 
50 kHz
NIDMM_VAL_AC_VOLTS
50
50e–6
50 V 
50 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
50e–6
5
50 V 
100 kHz
NIDMM_VAL_AC_VOLTS
50
50e–6
50 V 
100 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
50e–6
6
50 V 
300 kHz
NIDMM_VAL_AC_VOLTS
50
50e–6
50 V 
300 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
50
50e–6