7. Troubleshooting 
 
If you have trouble when using this product with connected to the emulator, refer to the following procedures. 
 
Table 3    Items to check when having trouble with this product used 
Type 
Problems 
Items to check 
Used without the user system. 
After turning off the power, verify that this product is connected to the user system, and 
then use it. 
Installed on a wrong position. 
Applied more than 5.5[V] voltage 
or reverse voltage to the power 
pin Vcc1/Vss1. 
Improper 
use 
Applied more than 10[V] of 
voltage or more than -10[V] of 
reverse voltage to each pin other 
than the power pins. 
This product and the emulator may have been damaged due to improper use of this 
product. Turn the power off immediately and disconnect this product and the emulator. 
Make a request for support or repair according to the procedure shown in Figure 4. 
Add the detailed information about the improper use to the column of Cause of fault in 
the Repair Request Sheet. 
Error was detected by the 
self-check of the emulator. 
Check to see if this product is disconnected to the emulator. When the product is 
connected to the emulator, after turning off the power and disconnecting it, execute the 
self-check of the emulator again. 
See the chapter of the Self-check in the user’s manual of the emulator. 
The self-check needs to be performed with the emulator disconnected from the 
product. 
LED of this product lights up. 
It indicates that abnormal voltage is being supplied. Turn the power off immediately 
and disconnect this product and the emulator. Check the connection between the 
emulator and the user system, the power supply voltage, and inadequateness or 
misuse of the user system. 
Error 
FUSE of this product is 
disconnected. 
This product is damaged due to the overload that exceeded the allowable limit of this 
product. The emulator may also have been damaged. Make a request for support or 
repair according to the procedure shown in Figure 4. 
 
 
When requesting for support or repair, please inform us of this product being used. If you request repair for the emulator used with 
this product, send us both the emulator and this product (We cannot accept any request for repair of this product). 
 
 
 
 
 
 
Unconnected
See the Emulator’s user’s 
m a n u a l   t o   r e q u e s t   f o r 
support/repair.
Was this product    
connected to the
emulator?
Disconnect this product and the 
user system from the emulator and 
execute the self-check of the 
emulator.
Connected
Make a request for support or repair. Be sure to inform us of this product being 
used with the emulator, and add the following information of (1)(2) to the column 
of Operating environment on the Repair Request Sheet, and that of (3)(4) to the 
column of Symptoms.
(1) [Optional product : R0E330850ACBP0]
(2) [Serial number: (Rot number of R0E330850ACBP0)]
(3) [Results of the self-check : (LED status or the error massage)]
(4) [FUSE status of R0E330850ACBP0:
 (Normal, RVC12 disconnected or RVS12 disconnected)]
Support /repair request
Complete
Complete
Check  t o  s e e  if the FUSE of this 
product is disconnected.
Note: If you request repair for the emulator used 
with this product, send us both the emulator and 
this product.
 
Figure 4    Support/repair request flow when using this product 