  The following commands can not be used during data storage. 
 
Query command for weighing data.   
 
Query command for stable weighing data. 
 
SI 
Query command for weighing data.   
 
SIR 
Query command for continuous weighing data. 
 
Setting the function table   
Parameter settings for each output mode are as follows: 
                            Item 
 
Mode 
Data output 
mode 
Auto print 
polarity, 
difference 
Data memory 
function 
Interval 
time 
Key mode 
prt 0, 4, 5 Not used 
data 2 
Auto print mode A 
prt 1 
data 2 
Auto print mode B 
prt 2 
ap-a 0-2 
ap-b 0-2 
data 2 
 
Not used 
Interval memory mode 
prt 3 
Not used 
data 2 
int 0-8 
 
Set each item, depending on the situation, as follows:   
No 
d-no 0 
No 
5-td 0 
Data number   
Yes 
d-no 1 
Time only 
5-td 1 
No  
5-id 0 
Date only 
5-td 2 
ID number 
Yes 
5-id 1 
 
Time and date 
Both 
5-td 3 
    With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets. 
 
Recalling the memory data 
Confirm that the “Data memory ( data )” parameter is set to “2”. 
1  Press and hold the PRINT key until    reCall is displayed, then release the key. 
2  Press the PRINT key to enter the memory recall mode. Recall the data in memory using the 
following keys 
RE-ZERO key 
To proceed to the next data set. 
MODE key 
To go back to the previous data set. 
PRINT key 
To transmit the current data using the RS-232C interface. 
With SAMPLE held down, press the CAL key 
 
To delete the data currently displayed.   
Note: Deleting the data will not increase the amount of data that can be stored. 
CAL key 
To exit the memory recall mode. 
3  Press the CAL key to return to the weighing mode.