A&D AD-4212B-101/102/201/301 用户手册
55
The following commands can not be used during data storage.
Q
Query command for weighing data.
S
Query command for stable weighing data.
SI
Query command for weighing data.
SIR
Query command for continuous weighing data.
Setting the function table
Parameter settings for each output mode are as follows:
Item
Mode
Mode
Data output
mode
mode
Auto print
polarity,
difference
polarity,
difference
Data memory
function
function
Interval
time
time
Key mode
prt 0, 4, 5 Not used
data 2
Auto print mode A
prt 1
data 2
Auto print mode B
prt 2
ap-a 0-2
ap-b 0-2
ap-b 0-2
data 2
Not used
Interval memory mode
prt 3
Not used
data 2
int 0-8
Set each item, depending on the situation, as follows:
No
d-no 0
No
5-td 0
Data number
Yes
d-no 1
Time only
5-td 1
No
5-id 0
Date only
5-td 2
ID number
Yes
5-id 1
Time and date
Both
5-td 3
With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets.
Recalling the memory data
Confirm that the “Data memory ( data )” parameter is set to “2”.
1 Press and hold the PRINT key until reCall is displayed, then release the key.
2 Press the PRINT key to enter the memory recall mode. Recall the data in memory using the
following keys
RE-ZERO key
To proceed to the next data set.
MODE key
To go back to the previous data set.
PRINT key
To transmit the current data using the RS-232C interface.
With SAMPLE held down, press the CAL key
To delete the data currently displayed.
Note: Deleting the data will not increase the amount of data that can be stored.
CAL key
To exit the memory recall mode.
3 Press the CAL key to return to the weighing mode.