14-3  Memory for Calibration and Calibration Test Data 
 Calibration data (when and how it is performed) and calibration test data can be stored in 
memory. 
  All the data in memory is available to be output at one time to a printer or personal computer.   
  Up to 50 data sets of the latest calibration or calibration test can be stored. When the memory 
capacity has been reached, “fUl” illuminates. 
 
Storing the calibration and calibration test data 
Note 
If “
Clr” appears blinking in the upper left of the display, delete the data in memory. 
Store the calibration and calibration test data as follows: 
1  Set the “Data memory ( data )” parameter to “3”. 
2  Set the “GLP output (info)” parameter to “1” or “2”. 
3  With the settings above, each time calibration or calibration test is performed, the data is stored 
automatically. 
 
Transmitting the memory data 
Confirm that the “Serial interface ( 5if )” parameters are set properly. For details, refer to          
“10. FUNCTION TABLE” and “17. CONNECTION TO PERIPHERAL EQUIPMENT” and also confirm 
that the “Data memory ( data )” parameter is set to “3”. 
1  Press and hold the PRINT key until out is displayed, then release the key. 
2  Press the PRINT key to display Sout no  with “no” blinking. 
3  Press the RE-ZERO key to display    out go  with “go” blinking. 
4  Press the PRINT key to transmit all memory data using the RS-232C interface. 
5  The balance displays  Clear  when all memory data is transmitted. Press the CAL key to 
return to the weighing mode. 
 
Deleting the memory data 
1  Press and hold the PRINT key until out is displayed, then release the key. 
2  Press the SAMPLE key to display    Clear . 
3  Press the PRINT key to display SClr no  with “no” blinking. 
4  Press the RE-ZERO key to display SClr go  with “go” blinking. 
5  Press the PRINT key to delete all data 
6  The balance displays    out    when all the data has been deleted. Press the CAL key to return 
to the weighing mode.