National Instruments 6023E 用户手册

下载
页码 136
Chapter 4
Signal Connections
4-8
ni.com
Analog Input Signal Overview
The analog input signals for these devices are ACH<0..15>, ASENSE, and
AIGND. Connection of these analog input signals to your device depends
on the type of input signal source and the configuration of the analog input
channels you are using. This section provides an overview of the different
types of signal sources and analog input configuration modes. More
specific signal connection information is provided in the
section.
Types of Signal Sources
When configuring the input channels and making signal connections,
you must first determine whether the signal sources are floating or
ground-referenced.
PFI4/GPCTR1_GATE
DIO
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
GPCTR1_OUT
DO
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
PFI5/UPDATE*
DIO
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
PFI6/WFTRIG
DIO
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
PFI7/STARTSCAN
DIO
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
PFI8/GPCTR0_SOURCE
DIO
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
PFI9/GPCTR0_GATE
DIO
Vcc +0.5
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
GPCTR0_OUT
DO
3.5 at (Vcc -0.4)
5 at 0.4
1.5
50 k
Ω pu
FREQ_OUT
DO
3.5 at (Vcc-0.4)
5 at 0.4
1.5
50 k
Ω pu
AI = Analog Input
DIO = Digital Input/Output
pu = pullup
AO = Analog Output
DO = Digital Output
Note: The tolerance on the 50 k
Ω pullup and pulldown resistors is very large. Actual value can range between 17 kΩ and
100 k
Ω.
Table 4-3. I/O Signal Summary (Continued)
Signal Name
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA
at V)
Rise
Time
(ns)
Bias