National Instruments 1128 用户手册

下载
页码 97
Chapter 2
Using the SCXI-1127/1128
2-22
ni.com
Figure 2-19.  SCXI-1332 Terminal Block
The SCXI-1127/1128 installed with the SCXI-1332 provides you with an 
× 8 matrix without any extra wiring except for hooking up your signals to 
the columns and rows of the matrix. Refer to 
 for 
information on using multiple SCXI-1127/SCXI-1332 systems to build 
matrices larger than 4 
× 8.
Figure 2-20 shows an example of an SCXI-1332 connected to an 
SCXI-1127/1128. The SCXI-1127/1128 is configured for 4 
× 8 matrix 
operation by the software. This is an example of a power supply test station. 
The purpose of this test station is to measure the output voltage of a power 
supply under various load conditions. The loads are connected to the first 
four columns, C0...C3. The power supply is connected to row 0 and is the 
DUT. A DMM is connected to row 1 and is used to measure the power 
1
Column Expansion Bus
2
Row Expansion Bus
3
Row Connections
4
Column Connections
5
Cable Strain-Relief
6
Safety Ground Solder Lug
6
3
5
4
2
1
1