National Instruments NI 5422 用户手册

下载
页码 36
NI 5422 Specifications
32
ni.com
Environment
NI PXI-5422 Environment
Note
To ensure that the NI PXI-5422 cools effectively, follow the guidelines in the 
Maintain Forced-Air Cooling Note to Users included in the NI 5422 kit. The NI PXI-5422 
is intended for indoor use only.
Table 14.  
Specifications
Value
Comments
Operating 
Temperature
0 ºC to +55 ºC in all NI PXI chassis except the following:
0 ºC to +45 ºC when installed in an NI PXI-101x or 
NI PXI-1000B chassis. (Meets IEC-60068-2-1 and 
IEC-60068-2-2.)
Storage 
Temperature
–25 ºC to +85 ºC. Meets IEC-60068-2-1 and 
IEC-60068-2-2.
Operating 
Relative 
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56.
Storage 
Relative 
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56.
Operating 
Shock
30 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test 
profile developed in accordance with MIL-PRF-28800F.
Spectral and jitter 
specifications 
could degrade.
Storage Shock
50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test 
profile developed in accordance with MIL-PRF-28800F.
Operating 
Vibration
5 Hz to 500 Hz, 0.31 g
rms
. Meets IEC-60068-2-64.
Spectral and jitter 
specifications 
could degrade.
Storage 
Vibration
5 Hz to 500 Hz, 2.46 g
rms
. Meets IEC-60068-2-64. Test 
profile exceeds requirements of MIL-PRF-28800F, Class B.
Altitude
2,000 m maximum (at 25 °C ambient temperature)
Pollution 
Degree
2