NI 5422 Specifications
Safety, Electromagnetic Compatibility, and 
CE Compliance
Table 15.  
Specification
Value
Comments
Safety
The NI 5422 meets the requirements of the following 
standards of safety for electrical equipment for 
measurement, control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1
• CAN/CSA-C22.2 No. 61010-1
For UL and 
other safety 
certifications, 
refer to the product 
label or to 
ni.com/
certification
search by model 
number or product 
line, and click the 
appropriate link in 
the Certification 
column.
Emissions
EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Immunity
EN 61326:1997 + A2:2001, Table 1
Up to 4 mVpp noise (about –44 dBm) may be present 
on the output during the conducted immunity test. Use 
of the product at levels below –44 dBm will result in 
self-recoverable errors.
Good screening (shielding) techniques must be 
employed throughout the data acquisition system.