Canon TX-20 Full Auto Tonometer 手册

下载
页码 100
28
3NamesofParts
DATA Screen
This screen shows measurements. 
6
5
4
3
2
1
8
9
10
7
Intraocular pressure 
 : Shows measurements from highest to 
lowest reliability in order 1 to 10. (Measurement 
number: Up to 10) 
 : Shows measurements from newest to 
oldest in order 1 to 10. (Measurement number: Up 
to 10) 
Eye being measured 
R: Right eye 
L: Left eye 
Level of received light 
Indicates the amount of light received from light 
reflected from cornea when measuring intraocular 
pressure. The number of lines increases as the 
amount of received light increases. 
If there is little received light, light reflected from 
the cornea may be obstructed by the eyelid or 
eyelashes. However, it is also affected by the 
corneal reflectivity of the patient. 
Low reliability 
This mark appears when the analysis of the results 
or reliability of the measurement is considered 
low. Measured values that have this mark are not 
included in the standard value calculation. 
5  
Intraocular pressure standard value 
Representative intraocular pressure value that was 
statistically derived from measured values. Measured 
values with a low reliability mark are not used. 
External device connection status 
Shows the status of external device connection. 
 : USB device is connected 
 : Error has occurred in USB device connection
 : LAN is connected 
 : RS-232C device is connected 
No display: No connection 
ID key 
Press to switch to the PATIENTID screen. 
MEASURE key 
Press to switch to the Measurement screen. 
MENU key 
Press to switch to the MENU screen. 
10  DELETE key 
Press to clear all the measurements. 
The measurement values are deleted from the 
instrument’s memory. 