(3) Precautions for transparent object measurement 
To measure a larger diameter transparent glass rod, the external diameter of a 
plastic object, or the width of a transparent sheet, arrange the measuring units with 
the support bases fitted so the scanning laser beams are opposing one another, and 
set the segment    to (1 + 5). If the laser scanning direction of each unit is reversed to 
the direction shown below, measurement will be disabled. 
 
 
 
 
 
Scanning 
direction
 
Reception unit
 
Emission unit 
 
Measuring 
unit 2
 
 
 
 
Set the measuring units 
with the support bases 
facing one another.
 
 
 
 
Measuring 
unit 1
 
 
 
 
 
Scanning 
direction
 
 
 
If two measuring units are set up with the support bases 
facing one another, a step L will be produced between the 
two units through inconformity of the optical axis of the two 
units, as shown at the right. The step values according to 
model are listed below. 
 
Measuring unit 
2 side
 
Optical axis
 
Side view 
 
 
Model Step 
LSM-501S 10mm 
LSM-503S 10mm 
LSM-506S 10mm 
LSM-512S 22mm 
Measuring unit 
1 side
 
LSM-516S 33mm 
 
 
 
No. 99MBC094A
 
4-10