NI USB-6008/6009 User Guide and Specifications
14
ni.com
The differential input mode can measure 
±20 V signals in the ±20 V range. 
However, the maximum voltage on any one pin is 
±10 V with respect to 
GND. For example, if AI 1 is 
+10 V and AI 5 is –10 V, then the 
measurement returned from the device is 
+20 V.
Figure 8.  Example of a Differential 20 V Measurement
Connecting a signal greater than 
±10 V on either pin results in a clipped 
output.
Figure 9.  Exceeding 
±
10 V on AI Returns Clipped Output