National Instruments 6008 用户手册

下载
页码 27
© National Instruments Corporation
13
NI USB-6008/6009 User Guide and Specifications
PGA
The progammable-gain amplifier provides input gains of 1, 2, 4, 5, 8,
10, 16, or 20 when configured for differential measurements and gain 
of 1 when configured for single-ended measurements. The PGA gain is 
automatically calculated based on the voltage range selected in the 
measurement application.
A/D Converter
The analog-to-digital converter (ADC) digitizes the AI signal by 
converting the analog voltage into a digital code.
AI FIFO
The NI USB-6008/6009 can perform both single and multiple A/D 
conversions of a fixed or infinite number of samples. A first-in-first-out 
(FIFO) buffer holds data during AI acquisitions to ensure that no data 
is lost.
Analog Input Modes
You can configure the AI channels on the NI USB-6008/6009 to take 
single-ended or differential measurements. Refer to Table 5 for more 
information about I/O connections for single-ended or differential 
measurements.
Connecting Differential Voltage Signals
For differential signals, connect the positive lead of the signal to the AI
terminal, and the negative lead to the AI– terminal.
Figure 7.  Connecting a Differential Voltage Signal
AI+
Voltage
Source
NI USB-6008/6009
AI–