National Instruments 5441 用户手册
© National Instruments Corporation
15
NI 5441 Specifications
Sample Rate Range and Resolution
Sample Clock
Source
Source
Sample Rate Range
Sample Rate Resolution
—
Divide-by-N
23.84 S/s to 100 MS/s
Settable to (100 MS/s) / N
(1
(1
≤ N ≤ 4,194,304)
High
Resolution
Resolution
10 S/s to 100 MS/s
1.06 µHz
CLK IN
200 kS/s to 105 MS/s
Resolution determined by
external clock source.
external clock source.
External Sample Clock duty
cycle tolerance 40% to 60%.
cycle tolerance 40% to 60%.
DDC CLK IN
10 S/s to 105 MS/s
PXI Star
Trigger
Trigger
10 S/s to 105 MS/s
PXI_Trig<0..7>
10 S/s to 20 MS/s
DAC Effective Sample Rate
Sample Rate
(MS/s)
DAC
Interpolation
Factor
Effective Sample
Rate
DAC Effective
Sample Rate =
(DAC
Interpolation
Factor)
Sample Rate =
(DAC
Interpolation
Factor)
×
(Sample Rate)
Refer to the
section for OSP
Interpolation.
Interpolation.
10 S/s to
105 MS/s
1 (Off)
10 S/s to
105 MS/s
12.5 MS/s to
105 MS/s
2
25 MS/s to
210 MS/s
10 MS/s to
100 MS/s
4
40 MS/s to
400 MS/s
10 MS/s to
50 MS/s
8
80 MS/s to
400 MS/s
Sample Clock Delay Range and Resolution
Sample Clock
Source
Source
Delay Adjustment Range
Delay Adjustment
Resolution
—
Divide-by-N
±1 sample clock period
<10 ps
High-
Resolution
Resolution
±1 sample clock period
Sample Clock
Period/16,384
Period/16,384
External (all)
0 ns to 7.6 ns
<15 ps
Specification
Value
Comments