National Instruments NI 5102 用户手册

下载
页码 87
Chapter 4
Hardware Overview
4-12
ni.com
Figure 4-9 shows the relevant timing signals for a typical pretriggered
acquisition. The illustration represents five pretrigger and five posttrigger
scans, and above-high-level analog triggering is used. See the
section later in this chapter for more information on analog
trigger types.
Figure 4-9. Pretrigger Acquisition
Start Trigger
Scan Clock
Scan Counter
Terminal Count
First Scan
First Posttrigger Scan
Stop Trigger
End of
Acquisition
= Discarded Points
= Pretrigger Points
= Posttrigger Points
Trigger Level
Points Discarded
Five Pretrigger, Five Posttrigger
Points Stored