Chapter 5    Using the SCXI-1125
5-4
ni.com
Making High-Voltage Measurements
Another common use of the SCXI-1125 is to make measurements up to 
1000 VDC. Making measurements beyond ± 5 V requires use of the an 
attenuator terminal block. The SCXI-1327 and SCXI-1313A terminal 
blocks have a selectable attenuator for choosing between no attenuation or 
100:1 attenuation, which allows you to use the SCXI-1125 with up to 
300 V
rms
 when using the SCXI-1327 and up to 150 V
rms
 when using the 
SCXI-1313A.
The TBX-1316 has a fixed 200:1 attenuation, which allows you to use the 
SCXI-1125 with up to 1000 VDC, Measurement Category I.
The SCXI-1327 and SCXI-1313A also include a cold-junction sensor so 
you can combine thermocouple measurements with high-voltage 
measurements. When making signal connections, or when working with 
high-voltage signals, refer to the terminal block installation guide. If you 
are using the SCXI-1125 to measure signals with attenuation on the 
terminal block, an external bias resistor is not needed, because a bias 
resistor is already used for achieving the attenuation.
Table 5-1 lists the extended ranges of gain possible with the SCXI-1327 
and SCXI-1313A. Table 5-2 lists the extended gain possible with the 
TBX-1316.
Table 5-1.  Extended Gain and Range Using the SCXI-1327 or SCXI-1313A 
Overall 
Gain
Input 
Range
SCXI-1125 
Gain
SCXI-1327 
Attenuation
SCXI-1313A 
Attenuation
0.01
± 300 V
1
100
0.02
± 250 V
2
100
0.002
± 150 V
2
100
100
0.05
± 100 V
5
100
100
0.1
± 50 V 
10
100
100
0.2
± 25 V 
20
100
100
0.5
± 10 V 
50
100
100
2.5
± 2 V 
250
100
100