National Instruments 1142 用户手册

下载
页码 86
Chapter 2
Connecting Signals
2-12
ni.com
The digital output signal is pin 26. Pin 26 is SER DAT OUT and is 
equivalent to the SCXIbus MISO serial data output line.
The digital I/O signals of the SCXI-1141/1142/1143 module correspond 
to the digital I/O lines of an E/M Series DAQ device. Table 2-4 lists the 
equivalencies.
The digital timing signals are pins 36 and 43:
Pin 36 is SCAN CLK, the signal used as a clock for the 
SCXI-1141/1142/1143 module multiplexer counter. The DAQ device 
pulses this signal at the end of each conversion if the module is in 
multiplexed mode.
Pin 43 is a reserved digital input.
Table 2-4.  SCXIbus to SCXI-1141/1142/1143 Module Rear Signal Connector 
to DAQ Device Pin Equivalencies
SCXIbus Line
SCXI-1141/1142/1143
Rear Signal Connector
E/M Series 
DAQ Device
MOSI
SER DAT IN
DIO0
D*/A
DAQ D*/A
DIO1
INTR*
SLOT 0 SEL*
DIO2
SPI CLK
SER CLK
EXT STROBE*
MISO
SER DAT OUT
DIO4
Note: An * means the signal is asserted low.