BAROSTIM NEO LEGACY REFERENCE GUIDE 
8-2
 
Main Screen 
The Main Screen of the application contains the Patient Identification, IPG Status, Therapy Settings, and 
Schedule windows.  This screen provides a view of all pertinent therapy related settings It also provides the 
Session Notes space to allow entry of any notes to be included in the Session Summary Report.  The 
Session Summary Report, which contains the final therapy parameters and all pertinent session information, 
can be generated by clicking the “Save Report…” button.  When the communication session is completed, 
the user can press “End Session…”  This returns the software to the Device Selection Screen. 
Patient Identification 
The software displays the patient identification along with IPG model and  serial numbers  in the Patient 
Identification window.  The patient identification information can be changed by clicking on the  “Edit…” 
button.  Both Patient Name and Patient ID can be stored. 
Note: For privacy reasons, Patient Name will not appear on saved reports. 
IPG Status 
The IPG Status window is always active and provides several pieces of information related to the current 
status of the IPG. 
Current Therapy Status 
The Therapy Status is a real-time display that shows information about therapy being delivered by 
the IPG.  “No Therapy” indicates that no therapy is being delivered. 
Quickly–Stopping All Applied Therapies 
In case of patient discomfort or safety concern, ensure a good telemetry signal exists and select the 
Stop  button, next to the symbol ,    in the software to immediately stop all therapies.  To start 
therapy output, press the Resume button, which is in the same screen location. 
Refer to the Emergency Personnel Information in section 12 of this Reference Guide for additional 
information regarding nonprogrammer-based suspension of therapies, including magnet use. 
Battery Life 
The software displays an estimate of the expected life remaining of the IPG battery, the date of the 
Recommended Replacement Time  (RRT), and  the current battery voltage.   The battery life 
estimation is based on the currently programmed permanent  therapy and schedule settings.  The 
device should be scheduled for replacement on or before the RRT to prevent loss of therapy.  If the 
RRT date has already passed, the Battery Life status box turns yellow and indicates “RRT Alert”.  
The battery voltage displayed may decrease during a communication session due to the high power 
requirements of telemetry as well as impacts of aggressive parameter settings used during testing 
of therapy efficacy. 
Lead Impedance 
The impedance of both leads is displayed in the Lead Impedance portion of IPG Status. Make an 
immediate lead impedance measurement by clicking on the 
↻  symbol  in Lead Impedance. The 
lead impedance values  provide a measure of lead integrity and can indicate whether therapy 
delivery is functioning properly.  Impedance measurement results from unused or plugged 
connector ports are not meaningful.