BAROSTIM NEO LEGACY REFERENCE GUIDE 
8-6
 
 
Figure 7: Example Schedule 
Click on the “Edit…” button in the Schedule window to set and adjust time of day when therapy will be 
delivered.  The Schedule contains a list of entries.  Each entry is defined by a time period and the Therapy 
assigned to  it.  “Off” should be selected if periods of no therapy are desired.  Click Save to program 
Schedule changes or “Cancel” to ignore changes. 
Note:  The IPG Schedule is in relation to the local time set by the programmer computer. Patients who 
travel to other time zones and have more than one therapy in their Schedule will experience therapy 
transitions based on the time zone in use at the time of IPG programming. If the patient travels to a different 
time zone, the IPG schedule does not change to the new time zone. Clinicians should take this behavior into 
consideration when programming multiple therapies.IPG Diagnostics. 
Use the “IPG  Diagnostics…”  button on the Main Screen to navigate to the Management Center  screen.  
This screen provides the following information: 
o
 
Versions of embedded firmware 
o
 
The date and time EOS was encountered (N/A if EOS has not occurred). 
o
 
The times of the measurements of the maximum and minimum battery voltages. 
For troubleshooting purposes, CVRx can request users to press the “Save IPG Diagnostics” button. 
Estimating Longevity of Implantable Pulse Generator 
The battery lifetime of the IPG is dependent on device therapy settings. For example, a 24-hour therapy 
with 20 Hz therapy frequency, bilateral output pathway, 500  µs pulse width, and 6.0mA pulse amplitude 
into a 600  Ohm load will yield estimated device longevity of approximately 20  months.  Follow-up 
frequency for each patient should be adjusted based on the longevity results of each follow-up such that the 
next follow-up occurs no later than half of the number of months to predicted RRT  when RRT is greater 
than 3 months away.  When RRT is less than 3 months away, it is recommended that replacement be 
scheduled on or before the RRT to prevent battery depletion. 
NOTE: The longevity estimates provided in this section give the time from the start of therapy to the RRT 
for the device. 
Table  2  contains the estimated device longevity for a variety of parameter sets. For these calculations, a 
500µs pulse width and single 24-hour therapy was assumed.