TROUBLESHOOTING
BAROSTIM NEO LEGACY REFERENCE GUIDE
13-1
1
1
3
3
T
T
R
R
O
O
U
U
B
B
L
L
E
E
S
S
H
H
O
O
O
O
T
T
I
I
N
N
G
G
CVRx Contact Information
CVRx, Inc.
9201 West Broadway Avenue, Suite 650
Minneapolis, MN 55445 USA
Phone: (763) 416-2840
Fax: (763) 416-2841
Email: engineering@cvrx.com
www.cvrx.com
Minneapolis, MN 55445 USA
Phone: (763) 416-2840
Fax: (763) 416-2841
Email: engineering@cvrx.com
www.cvrx.com
Programmer System and Warning messages
This section provides a list of System and Warning messages you may encounter while attempting to use the
CVRx Programmer Software Application:
CVRx Programmer Software Application:
As part of the troubleshooting activity save a diagnostics file. This file should be sent to CVRx for advanced
trouble shooting support. To save the diagnostics file, open the CVRx Launcher application, click on the
“Programmer Diagnostics…” button and select “Save PGM Diagnostics
trouble shooting support. To save the diagnostics file, open the CVRx Launcher application, click on the
“Programmer Diagnostics…” button and select “Save PGM Diagnostics
System/Warning Message
Possible Cause(s)
Troubleshooting
.
PGM005 - Programmer Not
Connected
Connected
PGM006 - Programmer
Connection Problem
Connection Problem
PGM007 - Programmer Failure
•
Poor USB connection
•
Damaged Programmer
Interface
Interface
•
Damaged Programmer
Interface USB cable
Interface USB cable
1.
Ensure Programmer Interface USB cable
is properly connected.
is properly connected.
2.
Ensure Programmer Interface USB cable
is not cut or otherwise damaged.
is not cut or otherwise damaged.
3.
Ensure green power indicator is lit on
Programmer Interface.
Programmer Interface.
4.
Disconnect, and then reconnect the USB
cable.
cable.
5.
If problem persists, exit and then restart
application
application
6.
If problem persists contact CVRx
IPG002 - Unable to Interrogate
IPG
IPG
IPG007 - Programming Request
Three question marks “???” is
displayed for a parameter or
lead impedance measurement
displayed for a parameter or
lead impedance measurement
•
Loss of communication.
1.
Ensure that the Programmer Interface is
placed in its upright position with no
obstructions between the Programmer
Interface and the IPG.
placed in its upright position with no
obstructions between the Programmer
Interface and the IPG.
2.
Ensure the Programmer Interface front or
back side label is facing the IPG.
back side label is facing the IPG.
3.
Ensure the distance from the Programmer
Interface and the IPG is no more than 2
meters.
Interface and the IPG is no more than 2
meters.
4.
If problem persists exit and then restart
application
application
5.
If problem still persists contact CVRx
If software stops responding to keyboard or pointer device input: the user may use task manager (accessed by pressing cntrl-alt-
delete) and close the application. The user can then restart the application. If that is unsuccessful, the programmer computer can
be restarted to recover functionality.
delete) and close the application. The user can then restart the application. If that is unsuccessful, the programmer computer can
be restarted to recover functionality.