TROUBLESHOOTING 
BAROSTIM NEO LEGACY REFERENCE GUIDE 
13-1
 
1
1
3
3
 
 
T
T
R
R
O
O
U
U
B
B
L
L
E
E
S
S
H
H
O
O
O
O
T
T
I
I
N
N
G
G
 
 
CVRx Contact Information 
CVRx, Inc. 
9201 West Broadway Avenue, Suite 650 
Minneapolis, MN 55445 USA  
Phone: (763) 416-2840 
Fax: (763) 416-2841 
Email: engineering@cvrx.com 
www.cvrx.com 
Programmer System and Warning messages 
This section provides a list of System and Warning messages you may encounter while attempting to use the 
CVRx Programmer Software Application: 
As part of the troubleshooting activity save  a diagnostics file.    This file should be sent to CVRx for advanced 
trouble shooting  support.  To save the diagnostics file,  open the CVRx Launcher application,  click on the 
“Programmer Diagnostics…” button and select “Save PGM Diagnostics 
System/Warning Message 
Possible Cause(s) 
Troubleshooting 
PGM005 - Programmer Not 
Connected 
PGM006 - Programmer 
Connection Problem 
PGM007 - Programmer Failure 
• 
Poor USB connection 
• 
Damaged Programmer 
Interface 
• 
Damaged Programmer 
Interface USB cable  
1. 
Ensure Programmer Interface USB cable 
is properly connected. 
2. 
Ensure Programmer Interface USB cable 
is not cut or otherwise damaged. 
3. 
Ensure green power indicator is lit on 
Programmer Interface. 
4. 
 Disconnect, and then reconnect the USB 
cable. 
5. 
If problem persists, exit and then restart 
application 
6. 
If problem persists contact CVRx 
IPG002 - Unable to Interrogate 
IPG 
 
IPG007 - Programming Request 
Three question marks “???” is 
displayed for a parameter or 
lead impedance measurement 
• 
Loss of communication. 
1. 
Ensure that the Programmer Interface is 
placed in its upright position with no 
obstructions between the Programmer 
Interface and the IPG. 
2. 
Ensure the Programmer Interface front or 
back side label is facing the IPG. 
3. 
Ensure the distance from the Programmer 
Interface and the IPG is no more than 2 
meters. 
4. 
If problem persists exit and then restart 
application 
5. 
If problem still persists contact CVRx 
If software stops responding to keyboard or pointer device input: the user may use task manager (accessed by pressing cntrl-alt-
delete) and close the application.  The user can then restart the application.  If that is unsuccessful, the programmer computer can 
be restarted to recover functionality.