Vieworks Co. Ltd. FXRD-1012NAW 用户手册
Rev.1.0
Page 122 of 151
VW40-152-009
VIVIX-S 1012N User Manual
Voltage
Information
N/A
N/A
Remain
Information
N/A
N/A
Wireless
Detection
Decision
Defective wireless module
Contact a service eingineer.
Connection
Decision
Inconsistent environment of the
wireless communication
wireless communication
Defective wireless module
Check obstacles and distance
between a detector and SCU.
between a detector and SCU.
Contact a service eingineer.
Sensor
Impact Sensor Decision
Defective shock sensor
Contact a service eingineer.
Temperature
Decision
Defective temperature sensor Contact a service eingineer.
AED
Decision
Defective AED sensor
Contact a service eingineer.
Memory
Detection
Decision
The calibration data is
inapplicable.
inapplicable.
Not available to save backup
images.
images.
Not available to save logs.
Contact a service eingineer.
File System
Decision
The calibration data is
inapplicable.
inapplicable.
Not available to save backup
images.
images.
Not available to save logs.
Contact a service eingineer.
Status
Information
N/A
N/A
IC
FPGA
Decision
Not available to take images
fomr the detector.
fomr the detector.
Contact a service eingineer.
Fuel Gauge
Decision
Not available to check the
remaining of a battery pack.
remaining of a battery pack.
Contact a service eingineer.
RTC
Decision
Errors in log acquisition time. Contact a service eingineer.