Vieworks Co. Ltd. FXRD-1012NAW 用户手册

下载
页码 151
 
 
Rev.1.0
 
Page 122 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
Voltage 
Information 
N/A 
N/A 
Remain 
Information 
N/A 
N/A 
Wireless 
Detection 
Decision 
Defective wireless module 
Contact a service eingineer. 
Connection 
Decision 
Inconsistent environment of the 
wireless communication 
Defective wireless module 
Check obstacles and distance 
between a detector and SCU. 
Contact a service eingineer. 
Sensor 
Impact Sensor  Decision 
Defective shock sensor 
Contact a service eingineer. 
Temperature 
Decision 
Defective temperature sensor  Contact a service eingineer. 
AED 
Decision 
Defective AED sensor 
Contact a service eingineer. 
Memory 
Detection 
Decision 
The calibration data is 
inapplicable. 
Not available to save backup 
images. 
Not available to save logs. 
Contact a service eingineer. 
File System 
Decision 
The calibration data is 
inapplicable. 
Not available to save backup 
images. 
Not available to save logs. 
Contact a service eingineer. 
Status 
Information 
N/A 
N/A 
IC 
FPGA 
Decision 
Not available to take images 
fomr the detector. 
Contact a service eingineer. 
Fuel Gauge 
Decision 
Not available to check the 
remaining of a battery pack. 
Contact a service eingineer. 
RTC 
Decision 
Errors in log acquisition time.  Contact a service eingineer.