Vieworks Co. Ltd. FXRD-1012NAW 用户手册

下载
页码 151
 
 
Rev.1.0
 
Page 124 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
7.3 
Cleaning and Disinfection 
After using the detector and peripheral equipments for examination, use germicidal disinfecting wipes or 
cloth with mild diluted disinfectant detergent to clean surfaces of the product. 
 
Recommended Detergent Foam 
 
Recommended disinfectant wipe 
 
Wip’ anios manufactured by Anios 
 
Sani-cloth Active Wipes by PDI 
 
Recommended disinfectant product 
 
Sulfa’safe 
 
Storage temperature: 5°C ~ 35°C 
 
How to Use Detergent Foam 
1  Prepare the disinfectant detergent and a clean and dry non-woven cloth. 
2  Use the spray bottle to spray detergent to the cloth and clean the equipment. 
3  After it has been cleaned, leave the equipment un-used for 15 minutes. 
4  Conduct cleaning once a week or in case of contamination. 
 
 
 
Do not re-use wipes. 
 
Be careful to use disinfectant detergent which can cause irritation to eyes and skin. 
 
Use in well-ventilated areas, and wear gloves at all times. 
 
Do not clean the equipment with its power on. 
 
Do not use abrasive brush and scraper to clean the product. 
 
Be careful not to make liquid soak when cleaning battery bay and the connector on the 
side of products. 
 
 
 
Other Disinfectant detergent compliant to conditions listed below may be used following 
proper procedures according to its own manual. 
 
European Biocidal Products designed for surface disinfection (Directive 98/8/EC) 
 
Detergent with composition of Didecyldimethylammonium chloride, polyhexamethylene 
biguamide hydrochloride.