Guidance and Manufacturer’s Declaration for EMC 
 
 
This device has been tested for EMI/EMC compliance, but interference can still occur in an 
electromagnetically noisy location. Attempt to maintain a suitable distance between electrical 
devices to prevent malfunction. 
 
9.4.1  Electromagnetic Emissions 
The Equipment Under Test (EUT) is intended for use in the electromagnetic environment specified below. 
The customer or user of the EUT should assure that it is used in such an environment. 
 
Immunity test 
Compliance  Electromagnetic Environment 
RF Emissions 
(CISPR 11) 
Group 1 
The EUT uses RF energy only for its internal function. Therefore, its 
RF emissions are very low and are not likely to cause any 
interference in nearby electronic equipment. 
RF Emissions 
(CISPR 11) 
Class B 
The EUT is suitable for use in all establishments other than domestic 
and those directly connected to the public low-voltage power 
supply network that supplies buildings used for domestic purposes. 
Harmonic emissions 
(IEC 61000-3-2) 
Class A 
Voltage fluctuations/ 
Flicker emissions 
(IEC 61000-3-3) 
Complies 
9.4.2  Electromagnetic Immunity 
The ViVIX-S 1012N system is intended for using in the electromagnetic environment specified below. The 
user of this system should assure that it is used in the following environment. 
 
Electrostatic Discharge (ESD) IEC 61000-4-2 
Item 
Description 
Immunity test 
 
Electrostatic discharge (ESD) IEC 61000-4-2 
IEC 60601 test condition 
 
Contact ±6㎸   
 
Air ±8㎸ 
Compliance Level 
 
Contact ±6㎸ 
 
Air ±8㎸ 
Electromagnetic Environment - 
Guidance 
 
Floors should be wood, concrete or ceramic tile. 
 
If floors are covered with synthetic material, the relative humidity 
should be at least 30%.