Vieworks Co. Ltd. FXRD-1012NAW 用户手册

下载
页码 151
 
 
Rev.1.0
 
Page 146 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
Electrical Fast Transient/Burst IEC 61000-4-4 
Item 
Description 
Immunity test 
 
Electrical fast transient/burst IEC 61000-4-4 
IEC 60601 test condition 
 
Power supply lines ±2㎸ 
 
Input / output lines ±1㎸ 
Compliance Level 
 
Power supply lines ±2㎸ 
 
Input / output lines ±1㎸ 
Electromagnetic Environment - 
Guidance 
 
Main power quality should be that of a typical commercial or hospital 
environment. 
 
Surge IEC 61000-4-5 
Item 
Description 
Immunity test 
 
Surge IEC 61000-4-5 
IEC 60601 test condition 
 
Differential mode ±1㎸ 
 
Common mode ±2㎸ 
Compliance Level 
 
Differential mode ±1㎸ 
 
Common mode ±2㎸ 
Electromagnetic Environment - 
Guidance 
 
Main power quality should be that of a typical commercial or hospital 
environment. 
 
Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11 
Item 
Description 
Immunity test 
 
Voltage dips, short interruptions and voltage variations on power 
supply input lines IEC 61000-4-11 
IEC 60601 test condition 
 
<5% Uт (>95% dip in Uт) for 0.5 cycle. 
 
40% Uт (60% dip in Uт) for 5 cycles. 
 
70% Uт (30% dip in Uт) for 25 cycles. 
 
<5% Uт (<95% dip in Uт) for 5 sec. 
Compliance Level 
 
<5% Uт (>95% dip in Uт) for 0.5 cycle. 
 
40% Uт (60% dip in Uт) for 5 cycles. 
 
70% Uт (30% dip in Uт) for 25 cycles. 
 
<5% Uт (<95% dip in Uт) for 5 sec. 
Electromagnetic Environment - 
Guidance 
 
Main power quality should be that of a typical commercial or hospital 
environment. 
 
If the user of the EUT image intensifier requires continued operation 
during power mains interruptions, it is recommended that the EUT 
image intensifier be powered from an uninterruptible power supply or 
a battery. 
 
 
Uт is the AC power prior to approving the test level voltage.