6.1.3  Defect Calibration Dialogue 
 
 
Vieworks provides the Defect Map data stored in the detector. However, you should 
download the data as the calibration cannot be performed in the detector automatically. 
 
This Defect Map data has been generated through stringent test in the production stages 
of detector. However, new defect may be newly formed while using the detector. In this 
case, make sure to calibrate defects in person and generate a new defect map. 
 
 
 
Defect Map 
Item 
Description 
Load Defect Map 
Loads the Defect Map calibration data located in the defect map path. 
Download Defect Map 
Downloads the Defect Map calibration data stored in the detector. 
Manual Defect Correction  Calibrates the defect manually. 
Defect Map Path 
Shows the path of applied Defect Map data.