Vieworks Co. Ltd. FXRD-1012NAW 用户手册
Rev.1.0
Page 99 of 151
VW40-152-009
VIVIX-S 1012N User Manual
6.1.4 Gain Calibration Dialogue
Vieworks provides the Gain data stored in the detector. However, you should download
the data as the calibration cannot be performed in the detector automatically.
the data as the calibration cannot be performed in the detector automatically.
This data is provided in case the Gain calibration cannot be performed. You should do
the Gain calibration in person as the detector condition can be different by the operating
method or use environment.
the Gain calibration in person as the detector condition can be different by the operating
method or use environment.
Gain Calibration
Item
Description
Target Value
Shows the target value of Gain calibration. (Recommended : 8000 or higher)
Current Value
Shows value after the exposure when processing the manual Gain calibration.
Stage
Selects the number of manual Gain calibration and shows each calibration stage.
Get
Starts the manual Gain calibration.
Cancel
Cancels the manual Gain calibration.
Load Gain
Loads the Gain calibration data located in the Gain path.
Upload Gain
Uploads the Gain calibration data to the detector for using portable mode.
Download Gain
Downloads the Gain calibration data stored in the detector.
Gain Path
Shows the path of applied Gain data.