VW40-152-009
 
VIVIX-S 1012N User Manual 
6.1.4  Gain Calibration Dialogue 
 
 
Vieworks provides the Gain data stored in the detector. However, you should download 
the data as the calibration cannot be performed in the detector automatically. 
 
This data is provided in case the Gain calibration cannot be performed. You should do 
the Gain calibration in person as the detector condition can be different by the operating 
method or use environment. 
 
 
 
Gain Calibration 
Item 
Description 
Target Value 
Shows the target value of Gain calibration. (Recommended : 8000 or higher) 
Current Value 
Shows value after the exposure when processing the manual Gain calibration. 
Stage 
Selects the number of manual Gain calibration and shows each calibration stage. 
Get 
Starts the manual Gain calibration. 
Cancel 
Cancels the manual Gain calibration. 
Load Gain 
Loads the Gain calibration data located in the Gain path. 
Upload Gain 
Uploads the Gain calibration data to the detector for using portable mode. 
Download Gain 
Downloads the Gain calibration data stored in the detector. 
Gain Path 
Shows the path of applied Gain data.