Stylus on-centre adjustment 
(if required)
Stylus on-centre adjustment
Stage 2    On-centre adjustment
7.
Each of the four screws A will move the
probe relative to the shank, in the X or Y
direction, as pressure is applied.
Tighten individually, backing off after each
movement.
8.
When the stylus tip run-out is less than
20 µm, fully tighten screws to 6 - 8 Nm
(4.4 - 5.9 lbf.ft).
9. For final centering, use screws A to move
the RMP60, progressively slackening on one
side and tightening the opposite screw, as
the final setting is approached, using two
hexagon keys.
Tip run out of 5 µm (0.0002 in) should be
achievable.
10. It is important that all four screws are tight
or tightened to 1.5 - 3.5 Nm (1.1 - 2.6 lbf.ft)
once the final setting has been achieved.
2.5 mm AF
4 mm AF
A
B
SHANK
PROBE