  ER920W Wireless Event Monitor  
 
22 
Service and Maintenance 
 
Cleaning  
Remove the batteries before cleaning the recorder.  Clean the battery 
terminals with a soft dry cloth.  Dampen a soft cloth with mild detergent 
and water to clean the recorder, lead wires, and belt clip. 
 
Remove any adhesives from the patient lead wires with an adhesive tape 
remover solution or swab.  Use a mild disinfectant.  Do not use alcohol 
or acetone on the lead wires since they could stiffen and the insulating 
plastic could crack. 
 
Service 
If there is a problem with the Monitor, review the problem descriptions 
and solutions listed on the next page.  If additional assistance is required 
contact customer support via phone, Fax or E-mail listed on the next 
page.  Call customer support before returning a Monitor to make 
shipping arrangements. 
A. 
Note there is not any preventative inspection or maintenance 
that can be performed by the end user. 
 
 
Service Items and Accessories 
 
Description Part 
Number 
Patient lead, 1 channel, Shielded 
350-0173-09 
Patient lead, 2 channel, Shielded 
350-0173-10 
Monitor belt clip / Holster 
100-1764-002 
Necklace 350-0180-00 
Operator manual 
600-0640-00 
3.6V Lithium Thionyl Battery Pack 
350-0294-00 
 
Notes: 
A.  Only accessories listed above are meant for use with the Monitor. 
B.  Dispose of batteries in accordance with Federal, State, and Local 
laws. 
 
 
 
 
 
 
  ER920W Wireless Event Monitor  
 
11
User Defined options continued. 
Display Option 
Description 
POST-EVENT 
SOUND 
 
Default = OFF 
ON = Sound on while recording 
during the Post-Event time.  
OFF = Sound off while recording 
an event. See Notes about sound 
for additional information. 
HOURLY 
REMINDER 
 
Default = ON 
ON = Ring every hour when an 
event is stored in memory. 
OFF = Do not remind patient an 
event is stored in memory. 
PREVIEW TIME 
 
Default = 30 sec 
The number of seconds the ECG 
signal is displayed for each 
channel when the patient cable is 
inserted.   
TRANSMIT 
SPEED 
 
Default = 1X 
The speed used for transmission 
of data to the receiving center.   
PACEMAKER 
DETECTION 
 
Default = ON 
Turn Pacemaker  stimulus 
detection ON/OFF.  
ARRHYTHMIA 
DETECTION 
 
Default = ON 
Turn Arrhythmia Detection 
ON/OFF.  When ON, The Brady 
Rate, Tachy Rate, and Pause 
Duration Threshold can be set. 
BRADY RATE 
 
Default = 45 bpm 
 
(Arrhythmia Detection must be 
on to access this function) 
TACHY RATE 
 
Default = 160 bpm 
 
(Arrhythmia Detection must be 
on to access this function) 
Continue on next page