FLIR Commercial Systems Inc. MO270X 用户手册
MO270-EU-EN-Temp02-11/11
13
Memory Logging Mode
Store a reading
To store a reading into one of the ten memory locations (Groups) press the F2-MODE soft-key
from the main display screen. Note that stored readings are date/time stamped.
Now use the F2-MODE soft-key again to select the Analog pointer or the Digital display mode.
Take a measurement and when the desired reading is displayed, press the F3-HOLD soft-key.
Press the F1-SAVE soft-key to begin the storing process. The memory locations screen will
appear.
Select a memory Group using the arrow buttons. Press the OK button when the desired memory
location is highlighted. The reading will now be stored in the selected memory group.
View stored readings
To view data from a memory group, press the F1-MENU soft-key from the main display screen
and navigate to the MEMORY sub-menu and press OK.
Highlight VIEW from the sub-menu and press OK. The groups list will appear.
Navigate to the desired group and press OK.
Use the up/down arrow buttons to scroll through the readings in the group. Use the CLEAR
button to delete a displayed reading.
Press F1-BACK to return to the Groups list or press EXIT to return to the normal operating mode.
Rename a Memory Group
To rename a memory group, press the F1-MENU soft-key from the main display screen and
navigate to the MEMORY sub-menu and press OK.
Highlight NAME from the sub-menu and press OK. The groups list will appear.
Navigate to the desired group and press OK.
An alpha-numerical screen will appear with the current group name shown at the top.
Use the arrow keys to select the desired digit to change and then press OK.
Now scroll to the new digit using the arrow keys. When the desired new digit is highlighted, press
OK and the old digit will be replaced by the new digit.
When editing is complete, press CONF to save entries and to return to the Groups list.
Deleting Memory
To delete stored data, press the F1-MENU soft-key from the main display screen and navigate to
the MEMORY sub-menu and press OK.
Highlight CLEAR from the sub-menu and press OK.
Three sub-menus will appear: INDIVIDUAL, GROUP, and ALL. Select INDIVIDUAL to delete one
reading from within one group; Select GROUP to delete an entire group; and select ALL to delete
all readings in all groups. Click OK to clear the item.