FLIR Commercial Systems Inc. MO270X 用户手册

下载
页码 16
MO270-EU-EN-Temp02-11/11 
14 
Calibration Verification 
  
Calibration Zero Check for Pinless Mode 
1. Switch the meter to the Pinless mode of operation (REL mode) from the
 F1-MENU
2. Ensure that the meter is not near any objects or surfaces. Hold the meter near the bottom to 
avoid contact with the pinless sensor. 
3. The display should read zero. 
4. If an error is displayed or if the meter reads other than zero, please return the unit for service. 
 
Calibration Check for Pin Mode 
1. Switch the meter to the Pin mode (%WME) from the
 F1-MENU
2. The three calibration check points are located in the holes at the top of the protective cap. 
3. The first measurement is made between the left test point and the center test point. The 
second measurement is made between the right test point and the center test point.  
4. The display should read 75 for the first measurement and 25 for the second measurement. 
5. If the readings are not correct, return the meter for service.