Section 6. Service and Troubleshooting
6.2.4
Diagnostics
The TankMaster WinSetup program lets you view the current device status. 
The View Diagnostic Register window shows a selection of database registers 
that gives you an instant view of how the gauge operates. You may also 
configure the window by adding registers of special interest.
To view and configure the diagnostic registers:
1. Select the 5900 gauge icon in the TankMaster WinSetup workspace.
2. Click the right mouse button and choose View Diagnostic Registers.
The register values in the diagnostics window are of read only type. They are 
loaded from the device as the window is opened.
A grey background color of the table cell in the Value column means that the 
register is of either Bitfield or ENUM type. An expanded Bitfield/ENUM 
window can be opened for this type of register. Double-click the cell to open 
the Expanded Bitfield/ENUM window.
If needed, the values can be presented as hexadecimal numbers. This 
applies to registers of the Bitfield and ENUM types. Select the Show in Hex 
check box to present Bitfield and ENUM registers as hexadecimal numbers.
The Configure button lets you open the Configure Diagnostic Registers 
window where you can change the list of registers to be displayed in the View 
Diagnostic Registers
 window. See the Raptor System Configuration Manual 
(Document No. 300510EN)
 for more information.
The Configure Diagnostic Registers window also has a Log Setup button for 
access to the Register Log Scheduling window which allows you to setup a 
log schedule for automatic start and stop of register logging. See “Logging Measurement Data” on page 6-12 for more information. 