MSI-7200 Dyna-Link   
   User Guide     Page 9
Note: To enter for example, 50 lbs on a 2000 lb Dyna-Link, the first two digits
should be entered as zeros (0050).
8) The next digit flashes. Use the UP, DOWN, and ENTER keys to finish
out the desired Set-Point value.
9) You are now back in the Set-Point Menu. Push UP or DOWN until the
display reads “End” and push ENTER (POWER) to return to normal
operation or use the UP/DOWN keys to select the other Set-Point for
modification.
Note: Set-Points are stored permanently in EEROM memory. Values are not
lost when the batteries are drained or removed.
SET-POINT Examples
Examples shown are for a 10000 lb Dyna-Link.
Example 1
Provides a warning when overload is close, and an additional second
warning when overload is reached.
SP1 = 9000, sense HI
SP2 = 10000, sense HI
Example 2
For in limit indication. When both Set-Points are on, the tension is
between 5000 and 5050 lbs.
SP1 = 5000, sense HI
SP2 = 5050, sense LO
With the optional relay box, the outputs could be combined for a single
in-limits indication.
Example 3
For an out of limit indication. When either Set-Point is on the tension is
outside the range of 2000 to 2500 lbs.
SP1 = 2000, sense LO
SP2 = 2500, sense HI
To Disable a SET-POINT
1) Turn the Dyna-Link off.
2) Hold down the PEAK key and turn the Dyna-Link on. After 1 second
release the PEAK key.
3) The display will read "Unit". Push UP (PEAK) or DOWN (ZERO) to
select the Set-Point (SP1 or SP2) to be disabled.
4) Push ENTER (POWER).
5) Use the UP/DOWN keys to select “OFF”. Push ENTER (POWER).