Measurement Systems International Inc DL7200RFL 用户手册
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
MSI-7200 Dyna-Link
•
User Guide Page 9
Note: To enter for example, 50 lbs on a 2000 lb Dyna-Link, the first two digits
should be entered as zeros (0050).
should be entered as zeros (0050).
8) The next digit flashes. Use the UP, DOWN, and ENTER keys to finish
out the desired Set-Point value.
9) You are now back in the Set-Point Menu. Push UP or DOWN until the
display reads “End” and push ENTER (POWER) to return to normal
operation or use the UP/DOWN keys to select the other Set-Point for
modification.
operation or use the UP/DOWN keys to select the other Set-Point for
modification.
Note: Set-Points are stored permanently in EEROM memory. Values are not
lost when the batteries are drained or removed.
lost when the batteries are drained or removed.
SET-POINT Examples
Examples shown are for a 10000 lb Dyna-Link.
Examples shown are for a 10000 lb Dyna-Link.
Example 1
Provides a warning when overload is close, and an additional second
warning when overload is reached.
SP1 = 9000, sense HI
SP2 = 10000, sense HI
warning when overload is reached.
SP1 = 9000, sense HI
SP2 = 10000, sense HI
Example 2
For in limit indication. When both Set-Points are on, the tension is
between 5000 and 5050 lbs.
SP1 = 5000, sense HI
SP2 = 5050, sense LO
With the optional relay box, the outputs could be combined for a single
in-limits indication.
between 5000 and 5050 lbs.
SP1 = 5000, sense HI
SP2 = 5050, sense LO
With the optional relay box, the outputs could be combined for a single
in-limits indication.
Example 3
For an out of limit indication. When either Set-Point is on the tension is
outside the range of 2000 to 2500 lbs.
SP1 = 2000, sense LO
SP2 = 2500, sense HI
outside the range of 2000 to 2500 lbs.
SP1 = 2000, sense LO
SP2 = 2500, sense HI
To Disable a SET-POINT
1) Turn the Dyna-Link off.
2) Hold down the PEAK key and turn the Dyna-Link on. After 1 second
2) Hold down the PEAK key and turn the Dyna-Link on. After 1 second
release the PEAK key.
3) The display will read "Unit". Push UP (PEAK) or DOWN (ZERO) to
select the Set-Point (SP1 or SP2) to be disabled.
4) Push ENTER (POWER).
5) Use the UP/DOWN keys to select “OFF”. Push ENTER (POWER).
5) Use the UP/DOWN keys to select “OFF”. Push ENTER (POWER).